Graseby™ 3500 Syringe Pump
Not made with Natural Rubber Latex
Non Sterile
Product Benefits
- Safe
- High Bolus Infusion Rate
- Occlusion Alarm
- AC Powered with backup battery operation. With more than 10 hours of life from full charge.
- Volume and Mass Totalizers
- Simple
- Preset (hands-free) and Manual (hands-on) Bolus
- Volume (mL/hr) and mass units Infustion Programming
- Optional TCI Capability:
- With the Astra Zeneca Diprifusor mode the 3500 pump can preform TCI ( Target Controlled Infusion) of Diprivan (Propofol)
- Makes Controlling the depth of anaesthesia as easy as using a vaporiser.
- effect site concentration avaliable at touch of a button
- Clear display of predicted blood concentration of Propofol
- Compensation for interrupted infusion.
- Time to ( Configurable) 'wake up'
- Time to target
- Smart
- Continuous Infusion and TCI Mode
- Accepts and automatically senses the size of syringes from the manufactures
- Communications Interface
- Pump History
- Bolus Hands free: programmed dose lets you concentrate on the patient not the pump. Hands on - for precise control in does/ respopnse situations. Bolus Volume dose and dose duration are calculated automatically.
Specifications
Reorder Code:
0138-0001
Details:
Infusion Modes
TCI and Continuous
Programming Modes
TCI target (blood) concentration, volume and mass units
Continuous Infusion Rate Range
0.1 - 400.0 ml/hr in 0.1 ml increments
Bolus Infusion Rate Range
0.1 - 1200.0 ml/hr in 0.1 ml increments
Mass Unit Programming Units
mg/hr, mg/kg/hr, mg/kg/min, mcg/hr, mcg/kg/hr, mcg/kg/min
Dipifusor™ Patient Details
Age: 16-100
Weight: 30-150kg
Target: 0.1 - 15 mcg/ml
Syringe Types
BD Plastipak 5, 10, 20, 30, 50/60 ml
Braun Omnifix 5, 10, 20, 30, 50/60 ml
Monoject 5, 10, 20, 30/35, 50/60 ml
Terumo 5, 10, 20, 30, 50/60 ml
Fresenius 50 ml
IMS Pumpjet 30 ml (prefilled)
Zeneca PFS 50 ml (1% and 2%)
Nipro 5, 10, 20, 30, 50/60 ml
JMS 5, 10, 20, 30, 50/60 ml
Drive Accuracy
The delivery volume is accurate to within ±2% over a complete syringe length (excluding the startup time)
Totalizers
TCI Mode
0 – 999.9 ml (in 0.1 ml increments)
0 – 99999.9 mg (in 0.1 mg increments)
Normal Mode
0 – 999.9 ml (in 0.1 ml increments)
0 – 99999.999 mg (in 0.001 mg increments)
History
Log of up to 1,500 events
Communications Interface
RS232 interface, communications protocol available on request
AC Power Supply
100-240V at 50/60Hz, 40VA
Battery Type
Sealed Lead Acid (Cyclon)
Battery Life
10 hours (fully charged and under normal operating conditions)
Electrical Safety Classification
Class II, Type CF
Fluid Ingress
IPX1 protection against drops of water falling vertically
Dimension
325 x 195 x 115 mm (with pole clamp fitted and plunger clamp closed)
Weight
Not exceeding 3.5 kg including batteries and pole clamp
Operating Conditions
+5 to +40°C, 30 to 75% RH, 700 to 1060 hPa
Design standards
IEC 601-1, IEC 601-1-2
Notes: Graseby 3500 TCI Syringe Pump incorporating the AstraZeneca DiprifusorTM system, UK variant.
Related Products