| Indications |
Intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. This pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia) |
| Pump Size |
1.6 in. x 4 in. x 5 in. excluding cassette or other accessories |
| Screen Size |
2.12 in. x 2.12 in. (320 pixels x 320 pixels) |
| Weight |
21 oz. including 4 AA Alkaline batteries, excluding other accessories |
| On-board Protocol Library |
Stores up to 500 protocols |
| Security |
Cassette/keypad lock & three customizable security access levels: keypad code; clinician code; administrator code |
| Accuracy |
+/- 6% (nominal) |
| Delivery Methods |
Continuous rate; PCA dose; Clinician bolus |
| Reservoir Volume |
0 to 9999 mL; programmable in 1 mL increments, displayed in 0.1 mL increments |
| Units |
Milliliters (mL), milligrams (mg), micrograms (mcg) |
| Concentration |
mg/mL:
0.1 to 0.5 mg/mL in increments of 0.1 mg/mL 0.5 to 1 mg/mL in increments of 0.5 mg/mL 1 to 15 mg/mL in increments of 1 mg/mL 15 to 100 mg/mL in increments of 5 mg/mL
mcg/mL:
1 to 15 mcg/mL in increments of 1 mcg/mL 15 to 100 mcg/mL in increments of 5 mcg/mL 100 to 500 mcg/mL in increments of 100 mcg/mL
|
| Continuous rate |
0 to 30 mL/hr (or the mg or mcg equivalent) |
| PCA Dose |
0 to 20 mL (or mg or mcg equivalent) Delivery rate (continuous rate + PCA dose): programmable from 40 to 175 mL/hr |
| PCA Dose Lockout |
1 minute to 24 hours in the following increments:
1 minute for values between 1 and 20 minutes 5 minutes for values between 20 minutes and 24 hours
|
| Max Doses |
Per Hour 1 to 60 |
| Delivery Limit Amount |
0.1 to 1000 mL (or the mg or mcg equivalent) in increments of:
0.01 mL from 0.01 to 0.5 mL 0.5 mL from 0.5 to 100 mL 1.0 mL from 100 to 1000 mL
|
| Given |
0 to 99,999.99 in 0.01 unit increments |
| Clinician Bolus |
0 to 20 mL (or mg or mcg equivalent). Delivery rate (continuous rate + clinician bolus): programmable from 40 to 175 mL/hr |
| Delivery Limit Method |
Delivery limit, max doses per hour or not in use |
| Delivery Limit Period |
1 to 12 hours in increments of 1 hour |
| Power Sources |
4 AA (IEC LR6) alkaline batteries; AC adapter; rechargeable battery pack |
| Battery Life (Alkaline) |
Approximately 120 hours at 10 mL/hr |
| Moisture Protection |
Splashproof (IPX4) per IEC 60529 |
| Event Log |
5000 events |
| Air Detector, integral |
Turn on or off |
| Air Detector sensitivity |
Low - Single bubble greater than 400 uL. Hight - Single bubble greater than 150 uL. |
| Alarms and Messages |
Multiple alarms, all color coded, many with option to “acknowledge” or “silence”. Alarms include high priority alarms; medium priority alarms; low priority alarms; informational messages and system fault alarms. |
| Alarm Volume |
High, medium, low |
| Alarm Sound Theme |
Standard, intense, distinctive |
| High Pressure Alarm |
18 +/- 9 psi |
| Graphs |
PCA dose graph; Delivery history and pie chart |
| Reports |
Given and PCA dose counters; delivery log; event log; protocol library summary; device information |
| Other |
Scroll keys; task-based soft key user interface; cassette latch; cassette/keypad lock; indicator lights; continuous backlight display with AC power; data interface port; remote dose cord; upstream and downstream occlusion sensors; polemount adapter plate; medication cassette reservoir; CADD®-Solis Lockboxes; compatible with CADD®-Solis medication safety software |